- Check the declaration of conformity
Legal basis
Medical Device Regulation (MDR) 2017/745 – Art. 19
Every medical device must have:
an EU Declaration of Conformity
✔ What you will find in it:
- the device class: I / IIa / IIb / III
- the intended purpose (medical vs aesthetic)
- the manufacturer’s details
This is the most important document for the buyer
- CE marking + notified body number
Art. 20 MDR
The device should bear:
- the conformity mark CE
- (for higher classes) the notified body number, e.g. CE 0123
✔ Interpretation:
- CE without a number → usually class I
- CE + number → class IIa, IIb, III
The higher the class → the higher the risk
- Instructions for use
Annex I MDR
The manufacturer must specify:
- indications
- contraindications
- risks
- possible complications
✔ What to look out for:
- whether burns / tissue damage are described
- whether the device acts at depth
- whether it requires special training
- Classification under the MDR (Annex VIII)
The buyer can ask:
“Under which MDR rule was the device classified?”
Most commonly for aesthetics:
- Rule 9 → energy-based devices
- Checking the intended purpose is crucial!
Is the device intended:
- for treatment (medical)
- to improve appearance (aesthetic)
✔ Why this matters:
- it affects the risk class
- it affects VAT
- it affects how the device is used
- The EUDAMED database (EU)
Art. 29 MDR
the official EU database of devices (still being partially implemented)
You can check:
- the manufacturer
- the class
- the certification
- Notified body certificate
For class IIa and higher:
the manufacturer MUST hold a certificate from a notified body
✔ What to check:
- the name of the body
- the number
- the validity
The most common mistakes buyers make
❌ “It has CE marking, so it’s safe”
👉 CE ≠ no risk
❌ “It’s medical, so it’s better”
👉 that depends on its intended purpose
❌ “Anyone can use it”
👉 the MDR does not regulate this
Practical interpretation for aesthetics
|
Technology |
Typical class |
Risk |
|
RF |
IIa / IIb |
medium |
|
HIFU |
IIb |
higher |
|
laser |
IIb |
higher |
The 3 most important questions to ask the seller
The buyer should always ask:
- What is the device’s MDR class?
- What is its intended purpose (medical vs aesthetic)?
- Under which MDR rule was it classified?
“A buyer can check a device’s level of risk on the basis of its MDR class, declaration of conformity and CE marking – however, CE marking alone does not mean low risk, only that safety requirements have been met.”